15
9.
When the temperature inside the ILR is observed to
be steady for several hours within +4° to +8°C only
then the vaccines to be preserved should be
transferred to the ILR.
The vaccines like DPT, which get damaged if frozen,
are adviced to be kept in the baskets provided, so that
they will not touch any of the colder surfaces on the
sides and the bottom. (fig. 25)
10.
After storing the vaccines keep the lid properly closed
and locked (fig. 26)
11.
The temperature of the bottom part of the ILR may be
observed on the separate dial thermometer placed on
it. If by any chance the temperature goes below 0°C
take out the bottom floor
remove the layer of frozen ice packs at the bottom
put a frest layer of not-frozen ice packs and replace
the floor
Now, the fresh ice packs will protect the vaccines from
freezing.
Adjust thermostat to a lower setting to obtain the
recommended temperature.
4. TO USE A REFRIGERATOR
Fig. 26.
Fig. 25.